You did what we told you to do. You bought a powered USB hub, plugged it in, moved your Dexis Platinum over to it, and now Windows calls it an “Unknown device.” Plug the sensor straight back into the computer and it works fine. So the hub is bad, right?

Probably not. When a Dexis sensor shows as an unknown device on a hub but works plugged in direct, the usual cause is that the Dexis driver went missing from Windows. The hub didn’t break anything. It just exposed a problem that was already there. The sensor is fine, the hub is fine, and you don’t need to box anything up. It takes about five minutes to fix.

Don’t want to dig through Device Manager?

Call us. One of our techs will connect remotely and put the driver back. Remote support is billed as a support session unless your account or purchase with us covers it, so ask when you call.

What Nick Found

Nick, one of our IT support techs, ran into this last week. The office was running a Dexis Platinum with VixWin on Windows 11, and they had just moved the sensor to the Plugable powered hub we recommend. Device Manager listed it under “Other devices” as Unknown device. Plugged straight into the PC, it took x-rays like nothing was wrong. The drivers looked up to date and properly installed, so updating them wasn’t going to help.

So Nick dug in. The Dexis driver package was gone from the place Windows keeps its drivers. Our best guess is that an old Windows update cleaned it out. We can’t prove that, but the driver was missing and nobody in the office had touched it.

And the Dexis driver is unsigned, which means Windows won’t install it on its own. You have to tell Windows you want it.

Why the Old Port Works and the Hub Doesn’t

Windows treats every USB port as its own door. The first time your sensor came in through a port, Windows set it up for that port and remembered it. Those ports keep working from that original setup, even after the driver files are gone.

Think of it like your house keys. The doors you’ve already opened are propped open. Every port on a new hub is a door you’ve never opened, and somebody threw the key away. Windows goes looking for the driver, finds nothing, and gives up with “Unknown device.”

That means any office whose driver is in this state will hit the problem the first time they move to a hub. Moving the sensor to a different USB port on the same computer will do it too. The hub just happens to be the first new door most offices open.

How to Tell If Your Dexis Unknown Device Is This Problem

You’re looking at the missing driver if all three of these are true:

  1. The sensor works when it’s plugged straight into the computer. Same sensor, same PC, same software.
  2. On the hub, Device Manager shows “Unknown device” under Other devices.
  3. The hardware ID matches. Right-click Unknown device, then Properties, then the Details tab, and pick Hardware Ids from the dropdown. If you see USB\VID_5328&PID_2009, this is your problem.

That PID_2009 is the sensor before its firmware loads. A working Platinum loads its firmware and shows up as PID_2010, “DEXIS Digital Intraoral Sensor.” Yours never gets that far because Windows has nothing to load.

The Fix, for Whoever Handles Your IT

This takes about five minutes and needs an administrator login on the computer. Plug the sensor into the hub first.

  1. Right-click Unknown device, then Update driver, then Browse my computer, then Let me pick from a list.
  2. Select Show All Devices, click Next, then click Have Disk.
  3. Click Browse. Don’t type the path in, or Windows will tell you the folder doesn’t contain driver software. On the PC Nick fixed, the file was at C:\Program Files (x86)\KaVo\Fusion\Win10\x64\DEXIS\. Select FusionDexis.inf, then Open, then OK.
  4. Pick Dexis Sensor before renumeration, click Next, and choose “Install this driver software anyway” when the warning comes up.
  5. The sensor should come back as DEXIS Digital Intraoral Sensor under Dental Intraoral X-ray Sensor. If a new Unknown device shows up instead, do it again with the same file and pick DEXIS Digital Intraoral Sensor.
  6. Take a test exposure.

Remember:

  • Don’t use oem59.inf, oem60.inf, FusionGX.inf or fusionkvo.inf. Those are Gendex and KaVo GXS-700 drivers. They list the Dexis IDs by name, but they can’t install them.
  • Can’t find FusionDexis.inf? Open a command prompt as administrator and run findstr /s /i /m /r "USB.VID_5328.PID_2009" C:\*.inf 2>nul to find it.
  • Getting a signature error or Code 52 instead of a warning? Try pnputil /add-driver "C:\Program Files (x86)\KaVo\Fusion\Win10\x64\DEXIS\FusionDexis.inf" /install from an administrator command prompt. If that fails too, get a signed Dexis Platinum driver from Dexis.
  • DO NOT turn off driver signature enforcement on a computer you take x-rays with. It fixes one driver and leaves the whole machine open.
  • Each new hub port needs this once. Pick a port, label it, and keep the sensor there.

Keep the Hub

Sure, unplugging the hub and going back to the old port makes the error go away. It also puts your sensor back on the same computer port that can’t always supply steady power, and that’s what kills sensors. A Dexis Platinum is especially sensitive to it. The hub stays. Fix the driver.

We’ve only confirmed this on the Dexis Platinum. The Gendex GXS-700 is built on the same hardware and also uses an unsigned driver, so it could do the same thing. We haven’t seen it yet. Whatever sensor you have, if it shows as unknown on a hub and works direct, check the hardware ID before you blame the hub.

And if your sensor shows up as unknown on every port, including plugged straight into the computer, that’s a different problem. Start with our sensor troubleshooting guide, check whether a failing USB port is the cause, and if it still won’t show up, send it to us for a free evaluation. If it isn’t worth fixing, we’ll tell you, and our pre-owned Dexis sensors come across the same bench as our repairs.

Sensor Still Not Showing Up?

If your Dexis sensor shows as an unknown device, drops its connection, or takes blank images, call us and we’ll remote in. If it turns out the sensor itself has failed, send it our way and we can probably fix it!

You can also talk to us on our online chat or email sales@sodiumdental.com.